Sensor Specifications
- Calibration: oxygen
- Flow range: 0 to 300 standard L/min
- Accuracy: ±2.0% of reading or 0.05 standard L/min @ 21 °C (Consult TSI for complete product specifications. Specifications are subject to change without notice.)
- Temperature range: 10°C – 50°C
- Pressure drop: 50 cm H2O at 200 L/min
- Response time: 3 msec (63% to full-scale flow)
- Output signals flow: 0 to 303 VDC non-linear, 2 channel
- Output signal temperature: 0 to 2.5 VDC non-linear
- Input voltage: 5 VDC ±10%, regulated
- Power consumption: 0.8 watts at full-scale flow
- Mating connector: Molex 87832-1020 (2 mm pitch, 2 row, 10 pin)
- Eeprom interface: I2C Communications
- Meets RoHS Directive 2011/65/EU
Features and benefits
- Fully calibrated and ready to install
- Mass flow measurements
- Fast response
- Accuracy as a percent of reading
- Low pressure drop